6.11 Inject function
This function performs an automatic injection movement. You can set the injection speed,
the injection time, the injection movement and fix the distance between the capillary and
the lower limit.
6.11.1 Inject menu and parameters
Abb. 6-2:Menu – Inject
Fig. 6-2: Menu – Inject
Parameter Value Range of values Increment Standard
Injection speed Set the injection speed in μm
per second
5 – 10000 5 300
Synchr. inject. Set the synchronized time of
injection
MOVE
IMMEDIATE
LIMIT
PRESSURE
LIMIT
Search+limit Change the lower limit and
the position of the capillary.
The distance between the
capillary and the limit
remains constant
OFF/ON OFF
Injection axial Switch the axial movement of
the function on or off
OFF/ON On
Parameter Range of values Function
Synchr. inject. MOVE Injection movement without injection pressure (when
using non-connected injectors)
IMMEDIATE Injection pressure starts with the injection movement
LIMIT Injection pressure at the end of the injection distance
(Z-axis Limit)
PRESSURE Injection pressure without injection movement
Zobrazit stránku 62