Operation
InjectMan
®
4
English (EN)
59
6.7.5 Overwriting a stored position
Prerequisites
A position is stored.
1. Press the softkey for a different position.
You will move to the position.
2. When the position has been reached, press and hold the softkey for the position to be
overwritten.
The old position is overwritten with the current coordinates.
3. Press the softkey.
An acoustic signal will sound.
In the softkey status field a filled circle is displayed.
The stored position is displayed in the coordinate field.
6.8 Vertical limits
For the Z-axis, a lower and upper limit can be defined. This prevents the capillary coming
into contact with the bottom of the Petri dish or moving against the condenser of the
microscope adapter.
•Lower limit Z-axis Limit
Upper limit – Upper limit
6.8.1 Defining the lower limit
Prerequisites
An application has been selected.
1. Position the capillary a little way above the slide.
2. Press Z-axis Limit.
The Z-coordinate is marked with L.
The value for the limit (Limit Z) is displayed.
Z-axis Limit is selected.
The capillary cannot be moved any lower.