FemtoJet
®
Operating Manual
eppendorf
®
is a registered trademarkPrinted in Germany
Your local distributor:
www.eppendorf.com/worldwide
Eppendorf AG
22331 Hamburg · Germany
Tel. +49 40 538 01-0
Fax +49 40 538 01-556
Eppendorf North America, Inc.
One Cantiague Road · P.O. Box 1019
Westbury, N.Y. 11590-0207 · USA
Tel. +1 516 334 7500
Toll free phone 800 645 3050
Fax +1 516 334 7506
E-Mail: info@eppendorf.com
Application Support
Europe, International:
Tel. +49 1803 666 789
E-Mail: suppor[email protected]
North America:
Tel. 800 645 3050 ext. 2258
E-Mail: suppor[email protected]
Asia, Pacific:
Tel. +603 8023 2769
E-Mail:
support_AsiaPacifi[email protected]
Titel_de_en.fm Seite 2 Montag, 4. Februar 2008 3:36 15
Zobrazit stránku 0
1 2 3 4 5 6 ... 56 57

Shrnutí obsahu

Strany 1 - Operating Manual

FemtoJet ® Operating Manual eppendorf ® is a registered trademarkPrinted in Germany Your local distributor: www.eppendorf.com/worldwide Eppendorf AG

Strany 2 - Eppendorf Offices

8 2 Device description 2.2 Short description Display Display of all relevant parameters and information. Function keysClean To apply the maximum pr

Strany 3 - Contents

9 2 Device description Variable regulator In the normal operating mode, the parameters for injection are set by the three variable regulators. The a

Strany 4

10 2 Device description 2.3 Startup Diagram of rear of device– Electrical connection is effected via the mains socket with integrated mains switch

Strany 5

11 2 Device description The drainage valve of the pressure cylinder is switched several times at the beginning of the initialization phase. The pres

Strany 6

12 3 Mode of operation 3.1 Display, variable regulators and keypad Before operating the device for the first time, users should read this section c

Strany 7

13 3 Mode of operation Slight deviations between the nominal value (specified value) and displayed compensation pressure p c (actual value) may occ

Strany 8

14 3 Mode of operation 3.1.3 Keypad Manual injection is performed for as long as the key is held down. Injection pressure pi is used. In the seco

Strany 9

15 3 Mode of operation Auto / Manu To switch between manual and automatic, time-controlled injection. In the case of manual injection, timeis coun

Strany 10 - 2 Device description

16 3 Mode of operation Standbyin the case of a brief interruption When key is pressed briefly: The device switches to the Standby mode without d

Strany 11

17 3 Mode of operation 3.2 Hand control / foot control The functions of the hand/foot control correspond to the functions of the and keys on the

Strany 12

Eppendorf Offices AUSTRALIA / NEW ZEALAND Eppendorf South Pacific Pty. Ltd.Tel. +61 2 98 89 50 00Fax +61 2 98 89 51 11E-mail: [email protected]

Strany 13

18 3 Mode of operation 3.3 Connecting the capillary holder Universal capillary holder The universal capillary holder is used to locate the injectio

Strany 14 - 3 Mode of operation

19 3 Mode of operation Other accessories Grip head 0, incl. 2 sets of O-rings and sealing washer.Using grip head 0 supplied, capillaries with an out

Strany 15

20 3 Mode of operation The usual values for the permanent compensation pressure P c are bet-ween 30 – 50 hPa (0,44 – 0,73 PSI) for Femtotips and si

Strany 16

21 3 Mode of operation 3.4.1 Compensation pressure p c The compensation pressure p c ensures that no medium flows into the capillary. Capillary

Strany 17

22 3 Mode of operation 3.4.3 Injection time t i Injection time t i gives the period for which injection pressure p i is maintained. t i is usua

Strany 18

233 Mode of operation3.4.5 Manual or automatic injectionIn the Manu operating mode, injection is performed for as long as the key or hand / foot co

Strany 19

243 Mode of operation3.5 The first injectionThe following are available: ● Adherent cells in a Petri dish (possibly on CELLocates to observe specific

Strany 20

253 Mode of operation3.6 Connecting an Eppendorf MicromanipulatorConnection to an Eppendorf Micromanipulator designed for automated injection is eff

Strany 21

263 Mode of operationFemtoJet (rear) with all connection options, with power unit 5171/InjectMan 51791 Foot control (optional) with Inject function2

Strany 22

273 Mode of operation3.7.1 Changing the capillaries3.7.2 Setting the pressure unit 3.7.3 Switching on/off the loudspeaker3.7.4 Switching on/off t

Strany 23

1 Safety precautions and application . . . . . . . . . . . . . . . . . . . 3 2 Device description . . . . . . . . . . . . . . . . . . . . . . .

Strany 24

283 Mode of operation3.7.5 Constant pressure3.7.6 Dewatering the pressure accumulatorFUNC 4This function can be used to apply a constant working pr

Strany 25

29 4.1 Cleaning Before cleaning the FemtoJet  , disconnect the plug.Ensure that no fluids enter into the inside of the injector in order to avoid sh

Strany 26

30 5 Troubleshooting 5.1 Potential errors – Pressure loss due to capillary not fitted with capillary holder connected.Solution: Connect the capillar

Strany 27

31 5 Troubleshooting key: Complete self-test– Checks compressor run time up to the point when a specific pressureis reached and thus for leaks in

Strany 28

32 5 Troubleshooting 5.2 Warnings As a rule, warnings are emitted if certain operating conditions are not (yet) reached or if an action is not poss

Strany 29

33 5 Troubleshooting 5.3 Error messages Error messages are emitted if certain operating conditions are not reached or the device records a deviatio

Strany 30

34 5 Troubleshooting Error 10 Supply pressure cannot be reachedThe compressor runs without the maximum supply pressure being reached.Possible cause:

Strany 31 - 4 Cleaning and Service

35 5 Troubleshooting Error 13 Supply pressure does not rise Possible cause: Technical defect Solution: Service Error 14 Massive leak in pressure acc

Strany 32 - 5 Troubleshooting

36 5 Troubleshooting Error 21 Injection pressure has not been reached Possible cause: Capillary holder not attached or poorly attached. (Broken) cap

Strany 33

37 5 Troubleshooting Error 25 No drop in working pressure Possible cause: Technical defect Solution: Service Error 26 Working pressure leak rate is

Strany 34

2 5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 305.1 Potential errors . . . . . . . . . . . . . . . . .

Strany 35

38 5 Troubleshooting Error 28 Working pressure leak rate is too highPossible cause: Capillary holder not attached or poorly attached. (Broken) capil

Strany 36

39 5 Troubleshooting Error 34 Valve 1 offset is too low Possible cause: Technical defect Solution: Service Error 35 Valve 1 offset is too high Possi

Strany 37

40Voltage/Frequency: 100 V – 240 V ± 10 %, 50 – 60 HzSet voltage: adapts automaticallyFuse for 100 – 240 V: T1.0 A, 250 V (2 pcs.)Power consumption:

Strany 38

417 Ordering informationOrder no. 5247 000.013 FemtoJet100 V – 240 V ± 10 %, 50 – 60 HzFully programmable microinjector with integrated pressure sup

Strany 39

427 Ordering informationOrder no. 5176 190.002 Universal capillary holder(incl. grip head 0, adapter for Femtotipsand spare O-rings)4910 000.018 Epp

Strany 40

437 Ordering informationOrder no.TransferMan NK 2Proportional micromanipulator for microinjection in suspension cells5188 000.012 Version Europe5188

Strany 41

447a Ordering information for North AmericaOrder no. FemtoJet100 V – 240 V ± 10 %, 50 – 60 HzFully programmable microinjector with integrated pressu

Strany 42 - 6 Technical data

457a Ordering information for North AmericaOrder no. 920007392 Universal capillary holder(incl. grip head 0, adapter for Femtotipsand spare O-rings)

Strany 43 - 7 Ordering information

467a Ordering information for North AmericaOrder no.TransferMan NK 2Proportional micromanipulator for microinjection in suspension cells920000011 Ve

Strany 44

47 A Adapter for Femtotips  18Application 3, 4 C Capillary change 27Clean 8pressure 20function 22Cleaning 29Commands 46Compressor 21, 23Compensatio

Strany 45

3 ● Observe operating instructions! ● The device is only designed for use in the research lab! ● The FemtoJet is a lab device. It may only be operate

Strany 46 - Accessories (not standard)

48 8 Index P Pressure– dewatering pressure accumulator 28– set pressure unit 26 Q Query 45 R Reset 9, 17function 16 S Safety precautions 3, 4Self

Strany 47 - Micromanipulation

49Appendix AOperation via a computerCertain commands and queries (Section 6.2) can be transmitted to the unit using the optional cable/PC interface. A

Strany 48

50Appendix AIf the command is faulty, an error message is returned instead of the answer:"ERR=eee"ERR: Error=: Separationeee: Error numberLo

Strany 49

51Appendix ACommand to enter parameters[C]Code[xxx]Input parameter [ssssssssss]Meaning of input parameterSet compensationpressure pc 043 0; 5 – 6,000

Strany 50 - 8 Index

52Appendix AError no.for incorrect operating constructionERR = [eee]Cause Solution 060 The first character is nota c or C.Check entry parameters. Chang

Strany 51 - Appendix A

53Appendix AOther error numbers as well as the error cause and solution are described in Section 5, "Troubleshooting". During remote control

Strany 52

54Appendix AA_Bedien_Computer_en.fm Seite 54 Mittwoch, 27. September 2006 5:26 17

Strany 53

A_Bedien_Computer_en.fm Seite 55 Mittwoch, 4. Juli 2007 2:07 14

Strany 54

4 ● Use the unit only for the described purpose! The unit and in particular the high pressure of the Clean function may not be used for purposes othe

Strany 55

5 2 Device description The FemtoJet  enables minute quantities of liquid to be injected into cells. The relevant parameters (injection pressure p

Strany 56

6 2 Device description 2.1 Overview 1 Display – Displays relevant parameters during injection or settingof parameters.– Guides operator through fun

Strany 57

7 2 Device description Standard accessories:1 Injection tube, 2 m1 Universal capillary holder with grip head for capillary diameters of 1.0 mm